Chapter 7 245
Post-Repair Procedures
Post-Repair Procedures
7. Post-Repair Procedures
Analog Base
Module (A0)
Perform the following required
adjustment.
- AUX Input
- Frequency Reference
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
Synthesizer
Module (A1)
Perform the following required
adjustment.
- Synthesizer Gain
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Receiver
Module (A2)
Perform the following required
adjustments.
- Source Output Power
- Receiver IF Range
- Receiver Port Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic accuracy test
- Uncorrected system performance
Table 7-1 Post-Repair Procedures for Analog Assemblies for Up to 8.5 GHz Options