246 Chapter 7
Post-Repair Procedures
Post-Repair Procedures
Level Vernier
Module (A3)
Perform the following required
adjustments.
- Source Output Power
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Distributer
Module (A4)
Perform the following required
adjustments.
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
Perform “To Execute the diagnostics Test” on
page 69
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Table 7-1 Post-Repair Procedures for Analog Assemblies for Up to 8.5 GHz Options
Replaced
Assembly or
Part
Required Adjustments
Correction Constants (CC)
Verification