Chapter 1 25
General Information
Required Equipment
1. General Information
Cable BNC(m)-BNC(m) Cable, 61 cm
Coaxial cable with Type-N (m)
connectors, 61 cm (24 in), 2 ea.
Agilent p/n 8120-1839
Agilent N6314A (p/n
8120-8862)
4
1
P,A
P,A ,T
Cable 20 inch 50 ohm cable Agilent p/n 8121-1834 1 P,T
Adapter N(m)-BNC(f) Adapter Agilent p/n 1250-0780 1 P,A
Torque Wrench Size: 3/4 inch
Torque: 136 N-cm
Agilent p/n 8710-1766 1 P,A
Personal Computer
with GPIB board
Windows XP, VEE7.5 or later 1 P,A
*1.P: Performance Tests, A: Adjustment, T: Troubleshooting
*2.The accuracy of a standard E4419A/B of ±0.02 dB is adequate for all tests/adjustment except “Dynamic Accuracy
Test”. This test requires a power meter with Option G12 or H12 that has been certified to a higer accuracy specifi-
cation. If an Option G12 or H12 power meter is not available, a test is provided on “Power Meter Accuracy Test”
on page 28 to verify the accuracy of a standard power meter.
Table 1-3 Alternative Test Equipment
Equipment Critical specifications Model Qty.
Use
*1
Frequency Counter Frequency: 50 MHz to 8.5 GHz
Accuracy: < 2.5 ppm
Agilent 53131/2A with Opt.010
and 124
1 P,A
Multimeter Agilent 34401A
Agilent 34410A
Agilent 34411A
1
1
1
P,T
P,T
P,T
Function Generator Agilent 33220A
Agilent 33250A
1
1
P,T
P,T
Power Sensor No Substitute Agilent E9304A with Opt. H19 1 P,A
Short Type-N(m) Agilent 11512A 4 P,T
Load Type-N(m) Agilent 909F Opt. 012 4 P
*1.P: Performance Tests, A: Adjustment, T: Troubleshooting
Table 1-2 Recommended Test Equipment
Equipment Critical specifications Recommended Model Qty.
Use
*1