26 Chapter 1
General Information
Required Equipment
2) Over 8.5 GHz Options
Table 1-4 lists the recommended equipment for performing maintenance on the E5071C.
Table 1-4 Recommended Test Equipment
Equipment Critical specifications Recommended Model Qty.
Use
*1
Frequency Counter 1 Frequency: 50 MHz to 8.5 GHz
Accuracy: < 2.5 ppm
Agilent 53181A with
Opt.010 & 124
1 P,A
Frequency Counter 2 Frequency: 50 MHz to 20 GHz
Accuracy: > 0.1 ppm
Agilent 53151A with
Opt.001
1 P
Frequency Standard Frequency: 10 MHz, Time Base
Error: ± 1× 10
–10
/year
Agilent 5071A 1 A
Power Meter No Substitute
Agilent E4419A/B
*2
1 P,A
Power Sensor No Substitute Agilent 8482A 1 P
Power Sensor No Substitute Agilent E4413A 1 P,A
Power Sensor No Substitute Agilent E9304A with Opt.
H18
1 A
Function Generator Agilent 33120A 1 P,A
Multimeter Agilent 3458A or 34401A 1 P,A
Dynamic Accuracy
Test Kit
No substitute Agilent Z5623A with Opt.
H01
1 P
Calibration Kit No Substitute Agilent 85052D 1 P,A ,T
Short 50ohm 3.5mm (f) Short Termination Agilent p/n 85052-60007 4 P, T
Short 50ohm 3.5mm (m) Short Termination Agilent p/n 85052-60006 1 P,T
Load Type 3.5mm (f) part of Agilent 85052D 2 P, A
Fixed attenuator (6
dB)
50 Ω, N(m)-N(f), VSWR 1.015 Agilent 8491A w/Opt.006
and H60
1 P
Fixed attenuator (10
dB)
50 Ω, N(m)-N(f) Agilent 8491A/B/C 1 T
Handler I/O Test Kit No substitute Agilent p/n E5070-65001 1 T
Low Pass Filter No substitute Agilent p/n 9135 -5852 1 P
Cable BNC(m)-BNC(m) Cable, 61 cm
Coaxial cable with Type-N (m)
connectors, 61 cm (24 in), 2 ea.
Agilent p/n 8120-1839
Agilent N6314A (p/n
8120-8862)
4
1
P,A
P,A ,T