48 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
2 Horizontal Controls
 
XY Time Mode
The XY time mode converts the oscilloscope from a volts- versus- time 
display to a volts-versus-volts display using two input channels. Channel 1 
is the X- axis input, channel 2 is the Y- axis input. You can use various 
transducers so the display could show strain versus displacement, flow 
versus pressure, volts versus current, or voltage versus frequency.
Example This exercise shows a common use of the XY display mode by measuring 
the phase difference between two signals of the same frequency with the 
Lissajous method.
1 Connect a sine wave signal to channel 1, and a sine wave signal of the 
same frequency but out of phase to channel 2.
2 Press the [AutoScale] key, press the [Horiz] key; then, press Time Mode 
and select "XY".
3 Center the signal on the display with the channel 1 and 2 position ( ) 
knobs. Use the channel 1 and 2 volts/div knobs and the channel 1 and 
2  Fine softkeys to expand the signal for convenient viewing.
The phase difference angle (θ) can be calculated using the following 
formula (assuming the amplitude is the same on both channels):
Figure 3 XY time mode signals, centered on display
A
 
B
 
D
 
C
Signal must
 
be cent ered in 
“X”
Measuring
phase difference
Signals 90 degrees
out of phase
Signals
in phase