Acquisition Control 11
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 157
See Also "Navigating Segments" on page 157
"Infinite Persistence with Segmented Memory" on page 157
"Segmented Memory Re-Arm Time" on page 158
"Saving Data from Segmented Memory" on page 158
Navigating Segments
1 Press the Current Seg softkey and turn the Entry knob to display the
desired segment along with a time tag indicating the time from the first
trigger event.
You can also navigate segments using the [Navigate] key and controls. See
"To navigate segments" on page 53.
Infinite Persistence with Segmented Memory
When data has been acquired to segmented memory, you can also turn on
infinite persistence (in the Display Menu) and press the Analyze Segments
softkey to create an infinite persistence display. The Analyze Segments
softkey appears when the acquisition is stopped and the segmented
memory feature is on.
Progress indicator
Sample rate

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 2000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 2000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 2000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals