Triggers 9
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 123
 
7 Select the qualifier time set softkey (< or >), then rotate the Entry 
knob to set the pulse width qualifier time.
The qualifiers can be set as follows:
• 2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz 
bandwidth models).
• 10 ns to 10 s for >< qualifier, with minimum difference of 5 ns 
between upper and lower settings.
Pulse width
trigger < qualifier
time set softkey
• When the less than (<) qualifier is selected, the Entry knob sets the 
oscilloscope to trigger on a pulse width less than the time value 
displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry knob sets the 
upper time range value.
Pulse width
trigger > qualifier
time set softkey
• When the greater than (>) qualifier is selected, the Entry knob sets the 
oscilloscope to trigger on a pulse width greater than the time value 
displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry knob sets the 
lower time range value.
Video Trigger
Video triggering can be used to capture the complicated waveforms of 
most standard analog video signals. The trigger circuitry detects the 
vertical and horizontal interval of the waveform and produces triggers 
based on the video trigger settings you have selected.
The oscilloscope's MegaZoom IV technology gives you bright, easily viewed 
displays of any part of the video waveform. Analysis of video waveforms is 
simplified by the oscilloscope's ability to trigger on any selected line of the 
video signal.