Measurements 13
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 185
Phase
Phase is the calculated phase shift from source 1 to source 2, expressed in
degrees. Negative phase shift values indicate that the rising edge of source
1 occurred after the rising edge of source 2.
1 Press the [Meas] key to display the Measurement Menu.
2 Press the Source softkey; then turn the Entry knob to select the first
analog channel source.
3 Press the Ty pe : softkey; then, turn the Entry knob to select Delay.
Phase =
Delay
Source 1 Period
× 360
Source 2
Source 1
Delay
Period