Getting Started 1
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 35
Front Panel Overlays for Different Languages
Front panel overlays, which have translations for the English front panel
keys and label text, are available in 10 languages. The appropriate overlay
is included when the localization option is chosen at time of purchase.
To install a front panel overlay:
1 Gently pull on the front panel knobs to remove them.
2 Insert the overlay's side tabs into the slots on the front panel.
19. USB Host port This port is for connecting USB mass storage devices or printers to the
oscilloscope.
Connect a USB compliant mass storage device (flash drive, disk drive,
etc.) to save or recall oscilloscope setup files and reference waveforms
or to save data and screen images. See Chapter 17, “Save/Recall
(Setups, Screens, Data),” starting on page 217.
To print, connect a USB compliant printer. For more information about
printing see Chapter 18, “Print (Screens),” starting on page 231.
You can also use the USB port to update the oscilloscope's system
software when updates are available.
You do not need to take special precautions before removing the USB
mass storage device from the oscilloscope (you do not need to "eject"
it). Simply unplug the USB mass storage device from the oscilloscope
when the file operation is complete.
CAUTION: Do not connect a host computer to the oscilloscope's
USB host port. Use the device port. A host computer sees the
oscilloscope as a device, so connect the host computer to the
oscilloscope's device port (on the rear panel). See “I/O Interface
Settings" on page 237.
There is a second USB host port on the back panel.
20. Digital channel
inputs
Connect the digital probe cable to this connector (MSO models only). See
Chapter 6, “Digital Channels,” starting on page 81.
21. Waveform
generator
output
Outputs sine, square, ramp, pulse, DC, or noise on the Gen Out BNC.
Press the [Wave Gen] key to set up the waveform generator. See
Chapter 16, “Waveform Generator,” starting on page 205.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 2000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 2000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 2000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals