205
Agilent InfiniiVision 2000 X-Series Oscilloscopes 
User's Guide
 
s1
16
Waveform Generator
To select generated waveform types and settings 205
To output the waveform generator sync pulse 208
To specify the expected output load 209
To use waveform generator logic presets 209
To add noise to the waveform generator output 210
To add modulation to the waveform generator output 210
To restore waveform generator defaults 215
A waveform generator is built into the oscilloscope. It is enabled by 
Option WGN or the DSOX2WAVEGEN upgrade. The waveform generator 
gives you an easy way to provide input signals when testing circuitry with 
the oscilloscope.
Waveform generator settings can be saved and recalled with oscilloscope 
setups. See Chapter 17, “Save/Recall (Setups, Screens, Data),” starting on 
page 217.
To select generated waveform types and settings
1 To access the Waveform Generator Menu and enable or disable the 
waveform generator output on the front panel Gen Out BNC, press the 
[Wave Gen] key.
When waveform generator output is enabled, the [Wave Gen] key is 
illuminated. When waveform generator output is disabled, the [Wave 
Gen] key is off.
The waveform generator output is always disabled when the instrument 
is first turned on.