Reference 21
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 287
CSV and ASCII XY files
"CSV and ASCII XY file structure" on page 288
"Minimum and Maximum Values in CSV Files" on page 288
File Header
12 bytes
Waveform Header 1
140 bytes
Waveform Data
Header 1
12 bytes
Voltage Data 1
buffer size
Waveform Header 2
140 bytes
Waveform Data
Header 2
12 bytes
Voltage Data 2
buffer size
Waveform Header N
140 bytes
Waveform Data
Header N
12 bytes
Voltage Data N
buffer size
Number of Waveforms = N = Number of Segments
Number of Waveform Buffers = 1
Index = 1
Time Tag = 0.0
Buffer Type = 1 (floating point)
Bytes per Point = 4
Number of Waveform Buffers = 1
Index = 2
Time Tag = time between segment 1 and 2
Buffer Type = 1 (floating point)
Bytes per Point = 4
Number of Waveform Buffers = 1
Index = N
Time Tag = time between segment 1 and N
Buffer Type = 1 (floating point)
Bytes per Point = 4