98 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
6 Digital Channels
 
you will discover the problem through random glitches or inconsistent 
data measurements. Use the analog channels to view ringing and 
perturbations.
Best Probing Practices
Because of the variables L, di, and dt, you may be unsure how much 
margin is available in your measurement setup. The following are 
guidelines for good probing practices:
• The ground lead from each digital channel group (D15–D8 and D7–D0) 
should be attached to the ground of the device under test if any 
channel within the group is being used for data capture.
• When capturing data in a noisy environment, every third digital channel 
probe's ground should be used in addition to the channel group's 
ground.
• High- speed timing measurements (rise time < 3 ns) should make use of 
each digital channel probe's own ground.
When designing a high- speed digital system, you should consider designing 
dedicated test ports that interface directly to the instrument's probe 
system. This will ease measurement setup and ensure a repeatable method 
for obtaining test data. The 01650- 61607 16-channel logic probe cable and 
the 01650- 63203 termination adapter are designed to make it easy to 
connect to industry- standard, 20-pin board connectors. The cable is a 2 m 
logic analyzer probe cable, and the termination adapter provides the 
proper RC networks in a very convenient package. These parts, as well as 
the 1251- 8106 20- pin, low- profile, straight board connector, can be 
ordered from Agilent Technologies.
To replace digital probe leads
If you need to remove a probe lead from the cable, insert a paper clip or 
other small pointed object into the side of the cable assembly, and push to 
release the latch while pulling out the probe lead.