Subsystem Command Reference 6
N9310A User’s Guide 113
LF Sweep State
:SWEep:LF:STATe ON|OFF|1|0
:SWEep:LF:STATe?
This command initiates your signal generator to sweep in a
different manner, according to the sweep trigger or point
trigger settings:
If the sweep trigger and point trigger are both set to
IMMediate
, sending the following command initiates an
LF sweep immediately:
“:SWEep:LF:STATe ON”
If either the sweep trigger or point trigger is set to KEY,
sending the command
“:SWEep:LF:STATe ON”arms
an LF sweep:
Sending command
“:TRIGger IMMediate” initiates
the LF sweep.
If either the sweep trigger or point trigger is set to EXT,
sending the commands “:SWEep:LF:STATe ON”arms
an LF sweep. No further command is needed to initiate
the LF sweep, but a proper external signal should be
input into the signal generator.
The query returns the current LF sweep state.
Range
On (1), OFF (0)
*RST value
OFF
Returned data
1, 0
Front panel access
Sweep > Sweep Mode > LF
Sweep > Sweep Mode > Off (abort an on-going LF
sweep)

Table of Contents

Other manuals for Agilent Technologies N9310A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9310A and is the answer not in the manual?

Agilent Technologies N9310A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9310A
CategoryTest Equipment
LanguageEnglish

Related product manuals