114 N9310A User’s Guide
6 Subsystem Command Reference
Amplitude Sweep State
:SWEep:AMPLitude:STATe ON|OFF|1|0
:SWEep:AMPLitude:STATe?
This command initiates your signal generator to sweep in a
different manner, according to the sweep trigger or point
trigger settings:
• If the sweep trigger and point trigger are set to
IMMediate
, sending the command
“:SWEep:AMPLitude:STATe ON”
initiates an amplitude sweep immediately:
• If either the sweep trigger or point trigger is set to KEY,
sending the command “:SWEep:AMPLitude:STATe
ON”
arms an amplitude sweep.
Sending command
“:TRIGger IMMediate”
initiates the amplitude sweep.
• If either the sweep trigger or point trigger is set to EXT,
sending the commands “:SWEep:AMPLitude:STATe
ON”
arms an amplitude sweep. No further command is
needed to initiate the
amplitude sweep, but a proper
external signal should be input into the signal generator.
The query returns the current amplitude sweep state.
Range
On (1), OFF (0)
*RST value
OFF
Returned data
1, 0
Front panel access
Sweep > Sweep Mode > Amplitude
Sweep > Sweep Mode > Off