EasyManuals Logo

Agilent Technologies N9310A User Manual

Agilent Technologies N9310A
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
118 N9310A User’s Guide
6 Subsystem Command Reference
Step Points
:SWEep:STEP:POINts <val>
:SWEep:STEP:POINts?
This command defines the number of points in a step sweep.
A step sweep must always have a minimum of two points
and can be configured to have as many as 1001 points.
The query returns the current number of points.
Step Dwell
:SWEep:STEP:DWELl <val> <unit>
:SWEep:STEP:DWELl?
This command sets the dwell time for each point in a sweep.
The query returns the current dwell time.
Range 2 to 1001
*RST value 10
Returned data format <val>
Front panel access Sweep > Step Sweep > # Points
Range 10 ms to 1s
Valid unit s, ms
Minimum increment 0.1 ms
*RST value 10.0 ms
Returned data format <val> <unit>
Front panel access Sweep > Step Sweep > Step Dwell

Table of Contents

Other manuals for Agilent Technologies N9310A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9310A and is the answer not in the manual?

Agilent Technologies N9310A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9310A
CategoryTest Equipment
LanguageEnglish

Related product manuals