118 N9310A User’s Guide
6 Subsystem Command Reference
Step Points
:SWEep:STEP:POINts <val>
:SWEep:STEP:POINts?
This command defines the number of points in a step sweep.
A step sweep must always have a minimum of two points
and can be configured to have as many as 1001 points.
The query returns the current number of points.
Step Dwell
:SWEep:STEP:DWELl <val> <unit>
:SWEep:STEP:DWELl?
This command sets the dwell time for each point in a sweep.
The query returns the current dwell time.
Range 2 to 1001
*RST value 10
Returned data format <val>
Front panel access Sweep > Step Sweep > # Points
Range 10 ms to 1s
Valid unit s, ms
Minimum increment 0.1 ms
*RST value 10.0 ms
Returned data format <val> <unit>
Front panel access Sweep > Step Sweep > Step Dwell