EasyManua.ls Logo

Allen-Bradley Dynamix 2500 - Page 140

Allen-Bradley Dynamix 2500
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
140 Rockwell Automation Publication 1441-UM001B-EN-P - September 2012
Chapter 5 Collecting and Reviewing Data
Alarms appear when there is an alarm condition, such as the removal of a
sensor on Channel X.
4. Once the reading is steady and the input signal is not overloading the data
collector, press Enter to accept the measured value.
Once the measurement is stored, you are taken back to the Route hierarchy.
You do not need to press Enter if you set Auto Store to On in the
Instrument Configuration screen.
The data collector automatically ranges and collects the
measurement.
See Instrument Configuration Parameters on page 59
.
TIP
The green status indicator illuminates when the measurement
is complete. See Status Indicators on page 22
.

Table of Contents

Related product manuals