EasyManua.ls Logo

Allen-Bradley Dynamix 2500 - Start Data Collection

Allen-Bradley Dynamix 2500
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rockwell Automation Publication 1441-UM001B-EN-P - September 2012 135
Collecting and Reviewing Data Chapter 5
Start Data Collection
After setting the data collector options and selecting a route, you are now ready to
start data collection.
Follow these steps to start collecting data.
1. Once you have selected a list, you begin collecting data from the designated
point on the route.
2. Connect the transducer and press Enter to start a measurement.
3. If necessary, press Enter again to collect a measurement depending on the
Auto Store setting in the Instrument Configuration screen.
See Instrument Configuration Parameters on page 59
.
You may notice that a Bias Check is taken if the parameter is set to Auto.
The Dynamix 2500 data collector supplies power to ICP
accelerometers through the A and B connectors. The ICP power `turns
on’ prior to collecting the first measurement point and remains on
until the following occurs:
You reach the end of the route.
The data collector advances to a measurement point that does not
require ICP power.
You exit the route menu.
Do not connect the channel to a BNC panel or non ICP device until the
Dynamix 2500 data collector advances to a non ICP measurement
point. Connecting the channel while collecting a measurement point
that requires ICP power may cause undesirable results, damage to the
equipment, or unexpected machinery trips.

Table of Contents

Related product manuals