106 Sequoia Series
This command is used to set the phase shift in eLoad Current Programming mode at the power-on.
Command Syntax [SOURce:] PONSetup:CURRent:PHASE
Parameters 0.1 to 5
Unit deg
Examples PONSetup:CURRent:PHASE
Query Syntax PONSetup:CURRent:PHASE?
Returned Parameters <NR2>
PONSetup:CURRent:PF
This command is used to set the power factor in eLoad Current Programming mode at the power-on.
Command Syntax [SOURce:] PONSetup:CURRent:PF
Parameters 0.1 to 5
Examples PONSetup:CURRent:PF
Query Syntax PONSetup:CURRent:PF?
Returned Parameters <NR2>
PONSetup:IMPedance
The impedance command Programs the Resistance, Inductance and Capacitance required as a load
for the UUT at the power-on.
Command Syntax [SOURce:]PONSetup:IMPedance
Parameters -1 to remove the connection, <NRf>
Example PONS:IMP -1 10 13
Since -1 is programmed to Resistance, It is considered as Resistance
is not connected.
Query Syntax [SOURce:]PONSetup:IMPedance?
Returned Parameters < NRf >
Related Commands SOUR:IMP?MIN, SOUR:IMP?MAX
PONSetup:OUTPut[:RELay]
This command sets the initial state of the output relay at the power on.
Command Syntax [SOURce:]PONSetup:OUTPut[:RELay <boolean>
Parameters 0|1|OFF|ON
Examples OUTP ON OUTP 0
Query Syntax PONS:OUTP?
Returned Parameters 0 | 1
PONSetup:PEAK:CURRent[:PROTection]
This command can be used to disable the peak current shutdown mode. It is factory disabled and
should be left disabled for most situations.