Sequoia Series 117
Command Syntax SOURCE: VOLTage:HIGH
Parameters <NR2>
Unit V (peak voltage)
Examples VOLT:HIGH 15
Query Syntax VOLT:HIGH?
Returned Parameters <NR2>
Related Commands [SOUR:]MODE VOLT:TRIG
VOLTage:LOW
This command sets the soft limits for minimum output voltage to which the unit could be programmed
Command Syntax SOURCE: VOLTage:LOW
Parameters <NR2>
Unit V (peak voltage)
Examples VOLT:LOW 5
Query Syntax VOLT:LOW?
Returned Parameters <NR2>
Related Commands [SOUR:]MODE VOLT:TRIG
VOLTage:MODE
Phase Selectable
This command determines how the output voltage is controlled during a triggered output transient.
The choices are:
FIXed The voltage is unaffected by a triggered output transient.
STEP The voltage is programmed to the value set by VOLTage:TRIGgered when a triggered
transient occurs.
PULSe The voltage is changed to the value set by VOLTage:TRIGgered for a duration
determined by the pulse commands.
LIST The voltage is controlled by the voltage list when a triggered transient occurs.
Command Syntax [SOURce:]VOLTage:MODE<mode>
Parameters FIXed | STEP | PULSe | LIST
*RST Value FIX
Examples VOLT:MODE LIST VOLT:MODE FIX
Query Syntax [SOURce:]VOLTage:MODE?
Returned Parameters <CRD>
Related Commands VOLT:TRG VOLT
VOLTage:RANGe[:LEVel]
This command sets the voltage range of the power source. On the Sequoia Series, there are two
voltage ranges available:
• 166 VAC or 333 VAC range for AC-mode and (AC+DC)-mode
• 220 VDC or 440 VDC range for DC-mode
The LIM:VOLT? Query may be used to determine which AC or DC voltage ranges are available to
select. Sending a parameter other than these will not be accepted.
Note that the query format of the VOLT:RANG command returns the voltage range value for the
selected range and mode. In AC or AC+DC mode, an AC voltage range value is returned. In DC mode,