46 Sequoia Series
Command Syntax CALibration:CURRent:LROFset <NRf>
Parameters <NRf> (value from 0 to 65535)
Examples CAL:CURR:LROF 35000
Query Syntax CAL:CURR:LROF?
Returned Parameters <NRf> (value from 0 to 65535)
CALibration :CURRent[:AC] :HROFset <NRf>
This command will result in the ADC calibration coefficient being calculated for the AC full scale output
current in the high range.
Command Syntax CALibration:CURRent:HROFset <NRf>
Parameters <NRf> (value from 0 to 65535)
Examples CAL:CURR:HROF 35000
Query Syntax CAL:CURR:HROF?
Returned Parameters <NRf> (value from 0 to 65535)
CALibration :CURRent:DC:LROFset <NRf>
This command will result in the ADC calibration coefficient being calculated for the DC full scale output
current in the low range.
Command Syntax CALibration:CURRent:DC:LROFset <NRf>
Parameters <NRf> (value from 0 to 65535)
Examples CAL:CURR:DC:LROF 35000
Query Syntax CAL:CURR:DC:LROF?
Returned Parameters <NRf> (value from 0 to 65535)
CALibration :CURRent:DC:HROFset <NRf>
This command will result in the ADC calibration coefficient being calculated for the DC full scale output
current in the high voltage range.
Command Syntax CALibration:CURRent:DC:HROFset <NRf>
Parameters <NRf> (value from 0 to 65535)
Examples CAL:CURR:DC:HROF 35000
Query Syntax CAL:CURR:DC:HROF?
Returned Parameters <NRf> (value from 0 to 65535)