EasyManuals Logo

Ametek SEQUOIA Series Programming Manual

Ametek SEQUOIA Series
231 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
Sequoia Series 69
Output Protection
OUTPut:PROTection:CLEar
This command clears the latch that disables the output when an overvoltage (OV), overcurrent (OC),
overtemperature (OT), or remote inhibit (RI) fault condition is detected. All conditions that generated
the fault must be removed before the latch can be cleared. The output is then restored to the state it
was in before the fault condition occurred.
Command Syntax OUTPut:PROTection:CLEar
Parameters None
Examples OUTP:PROT:CLE
Related Commands OUTP:PROT:DEL *RCL *SAV
4.10.2 Output Remote Inhibit Modes
OUTPut:RI[:LEVel]
This command sets the remote inhibit level mode. Factory default is LOW, which requires a contact
closure to open the output relay. The level can be reversed by setting it to HIGH. Once set, the RI
level setting is retained each time the power source is powered up.
Command Syntax OUTPut:RI:LEVel
Parameters LOW | HIGH
*RST Value LOW
Examples OUTP:RI:LEV HIGH
Query Syntax OUTP:RI:LEV?
Returned Parameters <CRD>
Related Commands OUTP
NOTE: When using the Remote Inhibit input, it will be necessary to disconnect any RI
connection to the Sequoia leader unit when turning on the Sequoia leader unit.
During initialization, the RI connection must be OPEN or initialization will be
halted with the message WAITING FOR AUXILIARY displayed on the LCD screen.

Table of Contents

Other manuals for Ametek SEQUOIA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek SEQUOIA Series and is the answer not in the manual?

Ametek SEQUOIA Series Specifications

General IconGeneral
BrandAmetek
ModelSEQUOIA Series
CategoryPortable Generator
LanguageEnglish

Related product manuals