64 Rev. 2
Remote Interface Reference
Command Set Reference : Instrument Configuration Queries
Command: DISPLAY:N2?
Function: Returns a 0 if the instrument is not configured to display
liquid nitrogen level on the home screen and a 1 if it is.
Returns: 0 or 1<CR><LF>
Default: N/A
Command: DISPLAY:HE?
Function: Returns a 0 if the instrument is not configured to display
liquid helium level on the home screen and a 1 if it is.
Returns: 0 or 1<CR><LF>
Default : N/A
Command: UNIT
Function: Returns the current liquid level units in use.
Returns: C, I, or %<CR><LF>
Default: %
Command:
N2:UNIT?
Function: Returns the current liquid nitrogen level units in use.
Returns: C, I, or P<CR><LF>
Default: P
Command:
RELAY1:CHannel?
Function: Returns a 0 if relay №1 is disabled, a 1 if the relay is assigned
to the nitrogen channel, and a 2 if the relay is assigned to the
helium channel.
Returns: 0, 1, or 2<CR><LF>
Default: 0
Command:
RELAY2:CHannel?
Function: Returns a 0 if relay №2 is disabled, a 1 if the relay is assigned
to the nitrogen channel, and a 2 if the relay is assigned to the
helium channel.
Returns: 0, 1, or 2<CR><LF>
Default: 0
Command:
FILL:CHannel?
Function: Returns a 0 if the auto fill relay is disabled, a 1 if the relay is
assigned to the nitrogen channel, and a 2 if the relay is
assigned to the helium channel.
Returns: 0, 1, or 2<CR><LF>
Default: 0