Rev. 2 65
Remote Interface Reference
Command Set Reference : Instrument Configuration Queries
Command: RELAY1:OPeration?
Function: Returns a 0 if relay 1 closes (alarms) when the level is
above the setpoint and a 1 if the relay closes (alarms) when
the relay is below the setpoint. By default, relay 1 is
configured as the high level relay with alarm condition when
level is greater than the setpoint.
Returns: 0 or 1<CR><LF>
Default: 0
Command:
RELAY2:OPeration?
Function: Returns a 0 if relay 2 closes (alarms) when the level is
below the setpoint and a 1 if the relay closes (alarms) when
the relay is above the setpoint. By default, relay 2 is
configured as the low level relay with alarm condition when
level is less than the setpoint.
Returns: 0 or 1<CR><LF>
Default: 0
Command:
HI
Function: Returns the HI setpoint limit (default: relay 1) in the
current units.
Returns: <value><CR><LF>
Default: 90%
Command:
LO
Function: Returns the LO setpoint limit (default: relay 2) in the
current units.
Returns: <value><CR><LF>
Default: 10%
Command:
RELAY1:SETpoint?
Function: Returns the Relay 1 setpoint in the current units
Returns: <value><CR><LF>
Default: 90%
Command:
RELAY2:SETpoint?
Function: Returns the Relay 2 setpoint in the current units
Returns: <value><CR><LF>
Default: 10%