66 Rev. 2
Remote Interface Reference
Command Set Reference : Instrument Configuration Queries
Command: A
Function: Returns the A setpoint limit (auto fill stop level) in the
current units
Returns: <value><CR><LF>
Default: 80%
Command:
B
Function: Returns the A setpoint limit (auto fill start level) in the
current units
Returns: <value><CR><LF>
Default: 20%
Command:
INTERVAL
Function: Returns the fill timer setting in minutes if the instrument is
configured for the nitrogen channel. Returns the sampling
interval in minutes if the instrument is configured for the
helium channel.
Returns: <value><CR><LF>
Default: 15
Command:
N2:INTERVAL?
Function: Returns the fill timer setting in minutes if the instrument is
configured for the nitrogen channel.
Returns: <value><CR><LF>
Default: 15
Command:
LENGTH
Function: Returns the sensor active length in the current units.
Returns: <value><CR><LF>
Default: N/A
Command:
SOURCE:REC_OUT?
Function: Returns a 0 if the 0-10 V
DC
Recorder Output is disabled, a 1 if
it is configured for the nitrogen channel, and a 2 if it is
configured for the helium channel.
Returns: 0, 1, or 2<CR><LF>
Default: 0