70 Rev. 2
Remote Interface Reference
Command Set Reference : Configuring Setpoints
Command: CONFigure:RELAY2:SETpoint <value>
Function: Configures the relay 2 trip setpoint in the current channel’s
units.
Returns: <CR><LF>
Default: 80
Command: HI=<value>
Function: Sets the HI setpoint in the current units (Relay 1).
Returns: <CR><LF>
Default: 80
Command: LO=<value>
Function: Sets the LO setpoint in the current units (Relay 2).
Returns: <CR><LF>
Default: 20
Command: A=<value>
Function: Sets the A setpoint (control band upper limit).
Returns: <CR><LF>
Default: 60
Note: The A setpoint (autofill stop) must be greater than the B
setpoint (fill start) and must also be between 0 and 100%.
Command: B=<value> or CONFigure:FILL:B <value>
Function: Sets the B setpoint (control band lower limit).
Returns: <CR><LF>
Default: 40
Note: The B setpoint (autofill start) must be less than the A
setpoint (fill stop) and must also be between 0 and 100%.
Command: INTERVAL=<value>
Function: Sets the liquid nitrogen fill time out value in minutes.
Returns: <CR><LF>
Default: 15
Note: Setting the value of INTERVAL to 0 disables the fill timer
function. The maximum value allowable is 600 minutes.
Command: CONFigure:INTerval:FILL <value>
Function: Sets the fill timer in minutes.
Returns: <CR><LF>
Default: 15
Note: Setting the value of FILL to 0 disables the fill timer function.