1313..22 CalibrationCalibration
MT1040A, MU104011A, MU104014A, and MU104015A do not require any
calibration in the meaning of adjustment after shipment from factory.
In order to ensure measurements made by these instruments within the
specifications, periodically running performance test (equivalent to the
inspections made when shipping) for the instrument. The recommended
frequency depends on the particular usage of the instruments, but generally
yearly performance test is highly advised.
The final decision on performance verification frequency is to be taken by the
user, as performance test verifies past measurements rather than guarantee
future measurements.
In the event of this equipment malfunctions, contact an Anritsu Service and
Sales office. Contact information can be found on the last page of the printed
or in a separate file of the PDF version, of this manual.
The separate file is "A4_address_book_e_******.pdf" stored in the Z2077A Utilities
ROM. "******" of the filename shows numeric characters.