EasyManua.ls Logo

Anthos A7 Plus - Intended Use; Classification and Reference Standards; Environmental Conditions; Transport and Packaging Conditions

Anthos A7 Plus
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A7 Plus OPERATOR'S MANUAL
EN
5
1.2. INTENDED USE
The dental units described in this manual are Medical Devices intended for dental treatment.
The dentist's board may hold up to 6 instruments.
The assistant’s board can hold 2 suction tubes and 3 instruments.
This equipment must be used only by adequately trained personnel (dentists and paramedics).
The device is intended for non-continuous operation with intermittent loads (see the operating times of the individual parts in the dedicated sections).
The device is classified as pollution degree 2.
Overvoltage class: II.
1.2.1. CLASSIFICATION AND REFERENCE STANDARDS
MEDICAL DEVICES classification
Classification of the dental unit in accordance with the indications given in annex IX of directive 93/42/EC as amended:
Class IIa.
ELECTRICAL MEDICAL EQUIPMENT classification
Classification of the dental unit in accordance with standard EN 60601-1 on safety of medical equipment: Class I - Type B.
Reference standards: the dental units described in this manual are designed and manufactured in compliance with IEC60601-1 3rd Ed. - 2007, IEC
60601-1-6 3rd Ed. - 2010, IEC 62366 1st Ed. - 2007, IEC 80601-2-60 1st Ed. - 2012, IEC 60601-1-2 3rd Ed., ISO 6875 3rd Ed. - 2011, ISO 7494-1
2nd Ed. - 2011 and EN 1717 (type AA and AB) standards as far as water mains safety devices are concerned.
Classification of RADIO DEVICES AND COMMUNICATION TERMINALS (only when the WIRELESS foot control is present)
Equipment classification according to Directive 99/05/EC Art.12: Class I.
1.2.2. ENVIRONMENTAL CONDITIONS
The equipment is to be installed in rooms that satisfy the following requirements:
temperature between 10 and 40 °C;
relative humidity between 30 and 75%;
atmospheric pressure ranging from 700 to 1060 hPa;
altitude ≤ 3000 m;
air pressure entering equipment ranging from 6 to 8 bar;
water hardness at the equipment inlet must not be above 25 °f (French degrees) or 14 °d (German degrees) for untreated drinking water. For water
with a higher hardness degree, it is recommended to soften water until it reaches a hardness degree between 15 and 25 °f (French degrees) or
between 8.4 and 14 °d (German degrees);
water pressure entering equipment ranging from 3 to 5 bar;
water temperature entering equipment not higher than 25 °C.
1.2.2.1. TRANSPORT AND PACKAGING CONDITIONS
Temperature between -10 and 70°C;
Relative humidity between 10 and 90%;
• Atmospheric pressure: from 500 to 1060hPa.
1.2.3. WARRANTY
CEFLA s.c. stands behind its products warranting safety, reliability and performance.
The warranty is valid only under the following terms:
Conditions given on the warranty certificate are observed.
Yearly scheduled maintenance is performed.
The equipment is used only as instructed in this manual.
The electrical wiring in the room in which the equipment is installed must conform to IEC 60364-7-710 (standards for electrical wiring in medical and
dental offices).
A 3x1.5 mm
2
line protected by a bi-polar cut-out that conforms to applicable standards (10 A, 250 V, distance between contacts at least 3 mm) must be
used to feed the equipment.
WARNING:
The colour of the three wires (POWER, NEUTRAL and EARTH) should satisfy the requirements of current standards.
Installation, repairs and, in general, any other operation requiring the casing to be opened are to be performed exclusively by personnel authorised by
ANTHOS.

Table of Contents