A7 Plus OPERATOR'S MANUAL
96
EN
12. DENTAL UNIT OPERATING UNIT MAINTENANCE PLAN
WHEN
PART
HOW
SEE PARAGRAPH
Before starting the
work day.
Drain condensate.
/
See paragraph 9.2.
CATTANI suction
separator.
Insert inside each suction filter a tablet of VF CONTROL PLUS.
See paragraph 9.5.
After each treatment.
Contra-angle handpiece.
Sterilise or disinfect outside.
See documentation
enclosed with
instrument.
Turbine.
Sterilise or disinfect outside.
See paragraph 5.4.
Micromotor.
Disinfect outside.
See paragraph 5.5.
Scaler.
Sterilise or disinfect outside.
See paragraph 5.6.
Syringe.
Sterilise or disinfect outside.
See paragraph 5.3.
Curing Light.
Sterilize the fibre optics, disinfect the outside.
See paragraph 5.7.
C-U2 camera.
Disinfect outside.
Do not use acid or harsh products.
See paragraph 5.8.
Surgical suction tubes.
Suck in about 1/2 litre of sanitising solution with each of the
suction tubes. Sterilize suction tube holder terminals.
See paragraph 9.4.
When needed.
Bowl.
Clean with off-the-shelf detergents formulated for ceramic
materials.
Do not use acid or harsh products.
See paragraph 7.1.
METASYS surgical
separator.
See documents enclosed with equipment.
/
DÜRR surgical separator.
See documents enclosed with equipment.
/
Operating light.
See documents enclosed with equipment.
/
Monitor on light pole.
See documents enclosed with equipment.
/
Removable instrument
cords.
Clean with suitable disinfectant in compliance with
manufacturer's recommendations.
Do not use acid or harsh products.
See paragraph 5.
Dental chair coated
surfaces and upholstery.
Clean with suitable disinfectant in compliance with
manufacturer's recommendations.
Spray product on disposable soft paper.
Do not use acid or harsh products.
See paragraph 1.4.
At the end of the
work day.
Bowl filter.
Clean filter in running water.
The content must be disposed of separately.
See paragraph 7.1.
Surgical suction filter.
Check filter and, in case suction capacity is reduced, change it
(code 97461845).
See paragraph 9.3.
Surgical suction tubes.
Clean the filter of the saliva ejector terminal.
See paragraph 9.4.
Hydraulic saliva ejector.
Clean the filter of the saliva ejector terminal.
See paragraph 6.6.
Weekly.
CATTANI surgical
separator.
Clean the separator container, drain valve and probes.
See paragraph 9.4.
Suction tube holder
terminals.
Lubricate the O-ring.
See paragraph 9.4.
Monthly.
Turbine return air filter.
Check the filter and replace it if necessary (code 97290014).
See paragraph 9.6.
Yearly.
Dental chair.
Contact the technical service department for general inspection.
/