5.5.5. RECIPROCATING OPERATING MODE
Characteristics.
- 2 selectable reduction ratios: 4:1 and 6:1,
- 3 selectable root canal drills,
- progressive alarm signal starting from 60% of maximum set torque,
Menu with micromotor extracted but not active.
All icon buttons are active and all available functions can be edited (see
paragraph 5.5.).
In addition to the standard settings, the RECIPROCATING mode also
allows adjusting the following functions:
Reversing micromotor drill rotation direction.
Select micromotor drill rotation direction by touching the relevant icon
button:
Standard rotation direction: rotation with reciprocating motion.
Reversed rotation direction: counter clockwise rotation with
continuous movement (helps releasing the drill from the root
canal).
Reversed rotation direction is signalled by a sound (3 BEEPS).
WARNING:
As soon as the micromotor is extracted, a sound (3 BEEPS)
warns the operator if the direction of rotation is reversed.
Operation at maximum set torque.
Rotation lock.
Rotation lock followed by rotation direction reversal.
List of root canal drills.
Press icon buttons or to scroll the list of pre-set root canal drills:
- Waveone
®
Gold
[1]
,
- Reciproc
®
[2]
,
- Reciproc
®
Blue
[2]
.
[1]
WAVE ONE
®
is a registered trademark of DENTSPLY SIRONA INC.,
York, Pennsylvania,USA.
[2]
RECIPROC
®
is a registered trademark of VDW GmbH, München,
Germany.
Certified contra angle pre-set list.
Press icon buttons or to scroll the list of the certified contra
angles:
Display text
Ratio
Reference contra angles
EVO E4
4:1
Goldspeed EVO E4
®
S6:1
6:1
Sirona Endo 6:1