A7 Plus OPERATOR'S MANUAL
EN
41
Removing the grip.
The nozzle ( a ) is screwed onto the grip ( b ).
Turn the selector switch counter-clockwise (LED g off) and press
button ( c ) to take the grip off the syringe casing.
Syringe removable cord.
Syringe features a removable cord to make cleaning operations easier
(see paragraph 5.).
Cleaning.
Use soft disposable paper towel dampened with
detergents/disinfectants.
WARNING:
Do not soak the syringe in disinfectant liquids or
detergents.
Recommended products: harsh products and/or products
containing acetone, chlorine and sodium hypochlorites.
Disinfection.
Syringe grip and spout: see paragraph 1.5.
NOTE:
bag before sterilising.
5.4. TURBINE
Connecting the handpiece and changing the drill.
Refer to the specific instructions provided with the handpiece.
Use.
WARNING:
Read the instructions for use of the various turbines.
Operating times: work 5 min., rest 5 min.
The cock ( f ) is used to adjust spray water quantity.
The cock ( e ) adjusts the amount of air spray for all the instruments.
Place the instrument in its working position.
NOTE:
instrument activation is highlighted by the relevant managing
screen appearing on the TOUCH DISPLAY.
The icon buttons available on the TOUCH DISPLAY are:
Increase of settable values.
Decrease of settable values.
Selection of turbine operating speed.
Fibre optics on/off.
Activation/deactivation independent water supply (with
SANASPRAY system, only).
Activation and selection of the type of spray delivered by the
instrument.
Main screen controls recall.
1% quick selection of turbine max. operating speed.
50% quick selection of turbine max. operating speed.
100% quick selection of turbine max. operating speed.
Activation/deactivation of
Peristaltic pump (if installed,
only).