A7 Plus OPERATOR'S MANUAL
EN
47
Press icon buttons or to change the quantity of saline solution
delivered by the peristaltic pump.
NOTE:
value setting range: 1 to 5. The quantity of solution
delivered associated to the settable values is:
- value 1: approx.35 cc/min,
- value 2: approx.50 cc/min,
- value 3: approx.70 cc/min,
- value 4: approx.90 cc/min,
- value 5: approx.100 cc/min.
NOTE:
the quantity of saline solution delivered by the peristaltic
pump can also be edited when the instrument is active.
Removable hose.
The micromotor features a removable hose to make cleaning operations
easier (see paragraph 5.).
Cleaning and maintenance.
Refer to the specific instructions provided with the instrument.
It is recommended to use Daily Oil (CEFLA s.c.) for lubrication.
WARNING:
Do not soak the instrument in disinfectant fluids or
detergents.
Inadvisable products: abrasive products and/or products
containing acetone, chlorine and sodium hypochlorite.
Sterilization.
Instrument handpiece only: see paragraph 1.5.
WARNING:
Carefully read the operating instructions supplied with the
instrument before attempting to sterilise.
Warnings for use.
WARNING:
Never put the contra angle on the micromotor while it is
running.
The drill release button must not be pressed during
operation!
Friction between the button and micromotor rotor
overheats the head and may cause burns.
The patient’s internal tissues (tongue, cheeks, lips, etc.)
must be protected against contact with the button by
using suitable instruments (mirrors, etc.).
The drills and various instruments attached to the
handpieces must comply with Biocompatibility Standard
ISO 10993.
5.5.1. RESTORATIVE OPERATING MODE
Characteristics.
- speed range: 100 to 40,000 Rpm (1:1 handpiece),
- speed range: 1 to 100%,
- reduction ratio customizable list,
- rotation speed editing mode, from fixed to variable and vice versa,
- alarm signal upon reaching the maximum set torque,
- quick reading of max. speed during motor rotation.
Menu with micromotor extracted but not active.
All icon buttons are active and all available functions can be edited (see
paragraph 5.5.).
NOTE:
any edited setting or value will be automatically saved inside
the selected working programme (ex. P1).