A7 Plus OPERATOR'S MANUAL
EN
51
5.5.3. IMPLANT OPERATING MODE
Characteristics.
- speed value adjustable from 5 to 2500 Rpm with value always referring
to drill, regardless of reduction ratio (reducers from 20:1 to 1000:1),
- torque adjusting range; 0.5 to 55.0 Ncm for certified reducers, or 1 to
100%,
- reduction ratio customizable list,
- alarm signal upon reaching the maximum set torque,
- setting button during motor rotation.
Menu with micromotor extracted but not active.
All icon buttons are active and all available functions can be edited (see
paragraph 5.5.).
NOTE:
any edited setting or value will be automatically saved inside
the selected working programme (ex. P1).
Hereinafter is the list of the acronyms of the types of certified contra
angles indicated on the TOUCH DISPLAY:
Display
text
Ratio
Display
torque
Torque
tolerance
to drill
Reference
contra angles
1000:1
1000:1
50 Ncm
±20%
All makes
256:1
256:1
50 Ncm
±20%
All makes
120:1
120:1
50 Ncm
±20%
All makes
ATR80I
80:1
70 Ncm
±10%
ATR ATR80I
®
ER64
64:1
55 Ncm
±10%
NSK SGM-
ER64i
®
ER32
32:1
55 Ncm
±10%
NSK SGM-
ER32i
®
K27:1
27:1
55 Ncm
±10%
Kavo IntraLux
CL09® +
Testina CL3
®
20:1
20:1
50 Ncm
±20%
All makes
75EKM
20:1
55 Ncm
±10%
W&H WI-
75E/KM
®
W&H
WS-75E/KM
®
R20L
20:1
55 Ncm
±10%
Castellini R20L
®
NSK X-SG20L
®
NSK S-Max
SG20
®
NSK
SGM ER20i
®
ATR20I
20:1
70 Ncm
±10%
ATR ATR20I
®
WS75
20:1
70 Ncm
±10%
W&H WS-75
®
W&H WI
75E/KM
®
CA20L
20:1
55 Ncm
±10%
Bien-Air
CA20:1L
®
16:1
16:1
±10%
All makes
K12:1
12:1
40 Ncm
±10%
Kavo IntraLux
CL04®+ Testina
CL3
®

Table of Contents