The handpiece cannot be put in autoclave; disinfect it on the outside with suitable products and
cover it with disposable plastic wrap.
Use soft disposable paper towels to disinfect the handpiece. Do not use harsh products or soak in liquids.
WARNING:
The curing light handpiece is NOT suitable for autoclave.
The curing light handpiece is not protected against penetration of liquids therefore it CANNOT be soaked in a solution to be
sterilized.
The outside of the lamp should be disinfected with the fiber optics on.
Do not use any type of disinfectant on the exposed optical surface of the handpiece when the fiber is removed. The surface will
become irreparably opaque if it comes into contact with disinfectant.
Maintenance.
This equipment does not require any particular type of maintenance.
Only technicians authorised by the manufacturer can replace and/or repair the handpiece and dental unit.
The handpiece has been purposely constructed in a manner that requires specific tools to open it and therefore it cannot be removed by the user.
The warranty is automatically void if the handpiece is altered in any way.
Troubleshooting.
When the lamp is removed, the light does not come on (no LED on control console illuminated).
Make sure the Midwest connection is correctly attached to the power cord.
Carefully screw the ring nut, try to put the lamp back in and then take it out again.
If the problem persists, contact technical service department.
Less light emitted.
- Make sure the fiber optics is not cracked or damaged in any way: replace it if it is so.
Contact the manufacturer to purchase original spare parts.
- Make sure there are no residues of curing products on the end of the fiber optics: if necessary, wipe off with alcohol or a plastic spatula.
If the handpiece must be sent back, please disinfect it.
Ship it back in its original packing.
In addition, please enclose a description of the fault with the shipping note.