A7 Plus OPERATOR'S MANUAL
EN
7
Quality of the water delivered by the dental unit.
The user is responsible for the quality of the water delivered by the dental unit and must adopt measures to maintain the water quality.
To ensure that delivered water is kept to quality standards, CEFLA s.c. recommends equipping the dental unit with an internal or external disinfection
system.
The dental unit, once installed, is exposed to possible contaminants coming from the water mains. So, to effectively overcome this problem, it is
advisable to install the dental unit only when its use will be daily and to perform the disinfecting procedures starting from the day in which it is installed
by following the instructions set forth in the relevant sections.
If the dental unit is equipped with the air separation device from water mains (EN 1717), make sure that the expected continuous supply of disinfectant
is also carried out by ensuring that the relevant tank contains a suitable quantity of disinfectant (see relevant paragraph).
NOTE:
contact your local dealer or Dentists Association for more detailed information about national laws and requirements.
Applied parts.
The parts of the equipment that come into contact with the patient while carrying out its functions correctly during standard use are: dental chair
upholstery, armrest, curing light fibre optics, syringe terminal, disposable camera protection, scaler bits, handpiece drills, suction tube terminals.
Non applied parts that may come into contact with the patient are: dental chair armrest support, dental chair lower casing, patient-side water unit
casing, water-to cup-spout, bowl, suction tubes, handpiece body.
WARNING:
Dental chair movement.
Make sure that the patient is ready to collaborate: ask
him/her to keep his/her hands and feet close to the body,
avoiding incorrect postures.
Check that the patient is sitting properly when moving the
dental chair (see figure).