Patient monitor User’s manual
Chapter 9-7
9.5.1 Setting SpO
2
Parameter
Select SpO
2
parameter area to enter SpO
2
parameter setting.
Setting NIBP measurement on the same limb
When NIBP and SpO
2
are simultaneously measured to the same limb of
patient, the operator shall enter the SpO
2
parameter setting menu and set
【NIBP Same Side】to【On】in order to ensure that the SpO
2
alarm status will
keep constant during the NIBP measurement until the completion of NIBP
measurement. If
【
NIBP Same Side
】
is set to
【
Off
】
, the low perfusion will
result in the SpO
2
measurement inaccuracy during the NIBP measurement and
trigger the physiological alarm of SpO
2
.
Select【QRS Volume】in the SpO
2
setting menu to change the QRS volume.
The volume can be set to a value from【1】to【10】, and also can be set to
Select
【
Wave Speed
】
in the SpO
2
wave setting menu, and select the wave
speed in the options as required.
The SpO
2
reading shown on the monitor is the average of data collected
within a specific time. The shorter the averaging time is, the quicker the
monitor responds to the change in the patient’s oxygen saturation level.
Contrarily, the longer the averaging time is, the slower the monitor responds
to the change in the patient’s oxygen saturation level, but the measurement
accuracy will be improved. When a critical patient is monitored, selecting
shorter averaging time will help understanding the patient’s state.
You can select【Average Time】in the 【SpO
2
Setup】menu, and select a
To set the PR source, you can select the【PR Setup】of PR parameter area,