Patient monitor User’s manual
Chapter 26-2
60601-1:1988 + A1:1991 + A2:1995:
Limit: type BF: for a.c.: 100μA, for d.c.: 10μA.
type CF: for a.c.: 10μA, for d.c.: 10μA.
⑨ Test the patient leakage current under single fault condition according IEC
60601-1:1988 + A1:1991 + A2:1995:
Limit: type BF: for a.c.: 500μA, for d.c.: 50μA.
type CF: for a.c.: 50μA, for d.c.: 50μA.
⑩ Test the patient leakage current Mains voltage on applied part: According
IEC 60601-1:1988 + A1:1991 + A2:1995:
Limit: type BF: for a.c.: 5mA.
type CF: for a.c.: 50uA
Warning: No use-serviceable parts inside, before servicing to
authorized representative or manufacturer.
Keep your equipment and accessories free of dust and dirt. To avoid
damage to the equipment, follow these rules:
1. Always dilute according the manufacturer’s instructions or use lowest
2. Do not immerse part of the equipment in the liquid.
3. Do not pour liquid onto the equipment or accessories.
4. Do not allow liquid to enter the case.
5. Never use abrasive materials (such as steel wool or silver polish), or erosive
cleaners (such as acetone or acetone-based cleaners).
Warning: Be sure to shut down the system and disconnect all power
cables from the outlets before cleaning the equipment.
Caution: If you spill liquid onto the equipment or accessories, contact
us or your service personnel.
26.2 Cleaning of the Monitor
Common detergent and non-corrosive disinfectant used in hospital can be
applied to clean monitor, however you must be aware that many kinds of