Patient monitor User’s manual
8
Long trend: 168h,minimum resolution is 1min(store
when power goes off)
High resolution trend: 2h, minimum resolution is 5s
128 groups of parameter alarm events and associated
parameter waveform at the alarm moment. Waveform
length is selected among 8s, 16s and 32s.
128 groups of ARR event and the associated waveform
for each waveform. The waveform length is selected
among 8s, 16s and 32s.
The storage time depends on the stored waveforms and
the quantity of them.
A.6 Measurement Specifications
EN 60601-2-27 / IEC 60601-2-27, ANSI/AAMI EC
13 , IEC 60601-2-25
3 lead: I, II, III
5 lead: I, II, III, aVR, aVL, aVF, Vx
12 lead: I, II, III, aVR, aVL, aVF, V1-V6
2.5 mm/mV(×0.25), 5 mm/mV(×0.5), 10 mm/mV
(
×1
)
, 20 mm/mV
(
×2
)
, 40 mm/mV
(
×4
)
, Auto
Diagnostic mode ≥ 90 dB
Monitor mode ≥ 105 dB
Surgery mode ≥ 105 dB
Diagnostic mode: 0.05 Hz to 150 Hz
Monitor mode: 0.5 Hz to 40 Hz
Surgery mode: 1 Hz to 25Hz
Diagnostic mode: 50Hz/60Hz AC filter shuts off