Patient monitor User’s manual
8
DDC signal
12C compatible
A.5.9 Data Storage
Trend data
Long trend: 168hminimum resolution is 1minstore
when power goes off
High resolution trend: 2h, minimum resolution is 5s
Parameter alarm
event
128 groups of parameter alarm events and associated
parameter waveform at the alarm moment. Waveform
length is selected among 8s, 16s and 32s.
ARR event
128 groups of ARR event and the associated waveform
for each waveform. The waveform length is selected
among 8s, 16s and 32s.
NIBP measurement
result
1000 groups
Holographic
waveform
The storage time depends on the stored waveforms and
the quantity of them.
A.6 Measurement Specifications
A.6.1 ECG
Standard
EN 60601-2-27 / IEC 60601-2-27, ANSI/AAMI EC
13 , IEC 60601-2-25
Lead type
3 lead: I, II, III
5 lead: I, II, III, aVR, aVL, aVF, Vx
12 lead: I, II, III, aVR, aVL, aVF, V1-V6
Lead standard
AHA, IEC
Gain
2.5 mm/mV×0.25, 5 mm/mV×0.5, 10 mm/mV
×1
, 20 mm/mV
×2
, 40 mm/mV
×4
, Auto
CMRR
Diagnostic mode 90 dB
Monitor mode 105 dB
Surgery mode 105 dB
Bandwidth (-3d B)
Diagnostic mode: 0.05 Hz to 150 Hz
Monitor mode: 0.5 Hz to 40 Hz
Surgery mode: 1 Hz to 25Hz
AC filter (50/60 Hz
Diagnostic mode: 50Hz/60Hz AC filter shuts off