Patient monitor User’s manual
Chapter 2-1
Before putting the system into operation, verify that the monitor,
connecting cables and accessories are in correct working order and
Please connect the monitor to a socket with protective earth. If the
socket does not have protective earth conductor, please do not use the
socket and use battery to provide power to the monitor.
To avoid explosion hazard, do not use the monitor in the presence of
flammable anesthetics or other flammable substance in combination
with air, oxygen-enriched environments, or nitrous oxide.
Do not open the monitor housings; electric shock hazard may exist.
All servicing and future upgrades must be carried out by the
personnel trained and authorized by manufacturer only.
When using the monitor with electrosurgical units (ESU), make sure
Do not come into contact with the patient during defibrillation.
Otherwise serious injury or death could result.
Do not rely exclusively on the audible alarm system for patient
monitoring. Adjustment of alarm volume to a low level or off may
result in a hazard to the patient. Remember that alarm settings should
be customized according to different patient situations and always
keeping the patient under close surveillance is the most reliable way
for safe patient monitoring.
The physiological data and alarm messages displayed on the monitor
are for reference only and cannot be directly used for diagnostic
To avoid inadvertent disconnection, route all cables in a way to
prevent a stumbling hazard. Wrap and secure excess cabling to avoid
risk of entanglement or strangulation by patient or personnel.