Patient monitor User’s manual
1
Appendix A Product Specifications
A.1 Safety Specifications
According to the MDD 93/42/EEC, the monitor is Type b equipment.
Classified according to the IEC60601-1 is as follows:
Parts
Classification
of protection
against
electric
shock
Degree of
protection
against
electric
shock
Degree of
protection
against
ingress of
liquid
Degree of
protection
against
hazards of
explosion
Mode of
operation
Mainframe
I
No mark
IPX1
Not suitable
Continuous
Secondary
display
No mark
EMS Module
NA
CF
MPS Module
Temp Module
IBP Module
ICG Module
SpO
2
Module
C.O. Module
BIS Module
CO
2
Module
BF
AG Module
CSM Module
Note:
I: Class I, internally and externally powered equipment.
When you doubt about the protecting earth integrality or protecting earth
lead of the equipment, youd better change the equipment to internally
powered equipment.
CF: Type CF applied part
BF: Type BF applied part