Patient monitor User’s manual
Chapter 12-4
Select IBP parameter area, and select【Unit】in the IBP setting menu, and
the options are
【
mmHg
】
,
【
kPa
】
or
【
cmH
2
O
】
.
Select 【 Display Format 】 in the IBP setting menu, and the options are
【S/D M】,【S/D】,【Mean】or【M S/D】.
Select【Wave Speed】in the IBP waveform setting menu, and select the
wave speed in the options as required.
Select【Scale】in the IBP waveform setting menu, and you can select【Auto】
or
【
Manual
】
. When selecting
【
Manual
】
, you can adjust the position of upper
scale and lower scale manually. While selecting 【 Auto 】 , the monitor will
adjust the wave scale according to actual IBP waveform range automatically.
Setting High Resolution Cursor
It is suggested to use the high resolution cursor when you want to scale the
IBP waveform more accuracy. Select【High Res. Cursor】in the IBP waveform
setting menu, and adjust the position of high resolution cursor as required.
The high resolution cursor will be displayed on IBP waveform at the same
12.6 Calculating Cerebral Perfusion Pressure
The monitor can calculate the difference between mean arterial pressure(ART)
and the intracranial pressure(ICP). The difference is cerebral perfusion
pressure , which is labeled CPP. Therefore, the CPP value will be displayed on
the screen only when the ART and ICP are displayed at the same time.