Patient monitor User’s manual
Chapter 11-5
Select NIBP parameter area, and select【Unit】 in the NIBP setting menu,
and The options are【mmHg】or【kPa】.
11.7.2 Setting Initial Cuff Inflation Pressure
Select NIBP parameter area, and select【 Inflation 】in the NIBP setting
menu, set the initial cuff inflation pressure according to patient type and
11.7.3 Setting Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure, and block the
venous blood vessel to assist venous puncture.
1. Select the NIBP parameter area, and enter the NIBP parameter setting
2. Select
【
Venipunc. Press
】
, and set to a proper value.
3. Select【Start Assist Venipunc.】to start it.
4. Puncture vein and draw blood sample.
5. Select
【
Start Assist Venipunc.
】
again to deflate the cuff. The cuff deflates
automatically after a set time if you do not deflate it.
Select NIBP parameter area, and select
【
Reset
】
, then the inflation value of
blood pressure pump restores to the initial value. In case the blood pressure
pump doesn’t work as normal but without any prompt, the blood pressure
pump can be checked by reset, thus the blood pressure pump in abnormal
condition due to unexpected reason will automatically restore.
The maintenance and calibration of NIBP measurement is necessary for
every two years. If you need to maintain NIBP, please contact the professional
service personnel. More details about NIBP maintenance, please refer to
Service Manual.