EasyManua.ls Logo

Anyview A6 - 18.5 BIS Setting

Anyview A6
305 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient monitor User’s manual
Chapter 18-3
18.5 BIS Setting
18.5.1 BIS Parameter Setting
Select BIS Smoothing Rate
Select Parameter Option. Open the BIS Setup menu. Select BIS
Smoothing Rate. There are 3 modes for selecting:10s,15s and30s.
Smoothed Rate decides how the average of BIS data is calculated by the
monitor. The smaller the value is, the more sensitive the response of the
monitor to changes in patients status will be. The larger the value is, the
smoother the BIS tendency will be, which indicates smaller changes and a less
impact of artifacts.
EEG Filter
Select Parameter Option. Open theBIS Setupmenu. SelectEEG Filter.
All filtering can be switched between ON and OFF by selecting the button
ON/OFF.
Trend Time
Select Parameter Option. Open theBIS Setup menu. Select Trend
Time
. There are 4 modes for selecting:
15min
,
30min
,
1h
and
2h. The function of power-off protection is supported.
Impedance Check
Select Parameter Option.Open the
BIS Setup
menu. Select
Impedance Check. Both Manual and Autoinspection modes can
be selected.
18.5.2 BIS Waveform Setting
Wave Scale
Select Parameter Option.Open the BIS Setup menu. Select Wave
Scale
. Five modes can be selected: 25uV/div, 50 uV/div, 100 uV/div, 250
uV/div and 500 uV/div;

Table of Contents

Related product manuals