d. Zeroing the Pressure Transducer
(
1
)
Turn off patient stopcock (3-way stopcock 2) before you start zeroing.
(2)The transducer must be vented to atmospheric pressure before zeroing.
( 3 ) The transducer should be placed at the same level with the patient
heart, approximately mid-axially line.
(4)Enter the main menu. Select【Module Setup】→【IBP1 Setup】 or
【IBP2 Setup】,select【Zero Cal.】to start zeroing.
( 5) Wait 3s for the zeroing procedure end and the pressure value that is
displayed on screen will approximately return to zero.
(6)After completing zero, close the stopcock to atmospheric pressure, and
open the stopcock to the patient.
Note: More details about zeroing the pressure tansducer information
and monitoring steps turn to Chapter 12: Section 12.7.
19.6 Alarm
Note: More details about alarm type, alarm level, alarm mode, state
icon and counter measure, turn to Chapter 5: section5.1、5.2、
19.6.1 Setting Alarm Volume
Enter the main menu, select【System Setup】→【Alarm Volume】, set its
value according to your need.
Caution: When 【 Alarm Volume 】 is set to “Close”and the alarm
conditions appear, the module won’t touch off sound
alarm. If you want the sound alarm, please be sure the
setup of【Alarm Volume】correct.