Patient monitor User’s manual
Chapter 12-2
6. Connect the extension tube of the transducer and blood vessel with the artery
needles and secure them, then make sure 3-way stopcock 1 and 3-way stopcock
2 (See the following figure) are in a state of ON. At this moment, BP waveforms
should appear on the screen of the monitor.
7. Flush the system once with saline water mixed with heparin every 15min,
this will keep the tubing without jam.
Warning:
The pressure measuring side of the transducer should be on the same
level as the heart of the patient in the process of zero-setting and
measurement.
If air bubbles appear in the tubing system, flush the system with the
infusion solution again. Air bubbles may lead to a wrong pressure
reading.
If measuring intracranial pressure with a sitting patient, level the
transducer with the top of the patients ear. Incorrect leveling may
give incorrect values.
Caution:The monitoring procedure may be different depending on
the different IBP accessories, please conduct the IBP
measurement as actual status.
3-way stopcock 2
Extension tube
3-way stopcock 1
Normal Saline flush
bag in pressure bag
Artery needle
Extension tube
Pressure sensor
Stopcock
Pressure transducer
Cable
Connecting to
IBP connector