EasyManua.ls Logo

Anyview A6 - Page 132

Anyview A6
305 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient monitor User’s manual
Chapter 14-3
vapor and eliminates the need of changing the adapter. It allows free positioning
of the AG sensor as well.
5. Unless the AG sensor is protected with an HME always position the AG
sensor with the indicating LED pointing upwards
Pre-use Check
1
Prior to connecting the AG airway adapter to the breathing circuit, verify
the O2 calibration by checking that the O2 reading on the monitor is correct
(21%). See Chapter 14.7.1 for instructions on how to perform room air
calibration.
2Always verify gas readings and waveforms on the monitor before
connecting the airway adapter to the patient circuit.
3
Perform the tightness check of the patient circuit with the AG sensor
snapped on the AG airway adapter.
The state of the LED on the AG sensor:
Indication
Status
Steady green light
System OK
Blinking green light
Zeroing in progress
Steady blue light
1)
Anesthetic agent present
Steady red light
Sensor error
Blinking red light
Check airway adapter
Note 1: Valid for IRMA multi-gas sensors only.

Table of Contents

Related product manuals