Patient monitor User’s manual
12
Excitation frequency
Sine wave: 64.8 kHz
Excitation current
0.3mA
Measuring impedance
range
0.2Ω to 3Ω
Baseline impedance
range
500 Ω to 2000Ω(using defibrillator proof cable with
resistance of 1kΩ)
Gain
×0.25, ×1
×2
×4
Bandwidth
0.25 Hz to 2.0Hz-3dB
Sweep speed
6.25mm/s, 12.5mm/s, 25mm/s
RR
Measurement range
0 rpm to 150 rpm
Resolution
1 rpm
Accuracy
±2 rpm or ±2% , whichever is the greater
Alarm range
0 rpm to 150 rpm, high/low limit can be adjusted
continuously.
Delay of apnea alarm
10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
A.6.3 NIBP
Standard
EN 60601-2-30 / IEC 60601-2-30, EN 1060-1, EN
1060-3, EN 1060-4, SP10
Measurement way
Automatic oscillometry
Measurement mode
Manual , Auto, STAT
Intervals for Auto
measurement time
1min, 2min, 3min, 4min, 5min, 10min, 15min, 30min,
60min, 90min, 2h, 4h, 8h, 12h
STAT mode cycle
time
Keep 5 minutes, at 5 seconds interval.
Measurement kinds
Sys, Dia, Map
Measurement range
mmHg
Sys:
Adult
Pediatric
30 to 270
30 to 235