Patient monitor User’s manual
15
Measurement protection
time
Initial inflation pressure
Adult: 120~280mmHg,default 160mmHg
Child: 80~170mmHg,default 120mmHg
Neonate: 60~140mmHg,default 90mmHg
Startup Initialization
Period
Intervals for AUTO
measurement time
1,2,3,4,5,10,15, 30, 60,90 minutes
2,4,8,12 hours
Overpressure Protection
Adult
Child
Neonatal
Hardware and software double protections
<300 mmHg
<300 mmHg
<150 mmHg
70% to 100%:±2%
0%~69%:unspecified
Normal:8s, slow: 16s,fast: 4s