15
Measurement protection
time
Adult: <180s
Child: <180s
Neonate: <90s
Initial inflation pressure
Adult: 120~280mmHgdefault 160mmHg
Child: 80~170mmHgdefault 120mmHg
Neonate: 60~140mmHgdefault 90mmHg
Startup Initialization
Period
7 seconds
Intervals for AUTO
measurement time
1,2,3,4,5,10,15, 30, 60,90 minutes
2,4,8,12 hours
Overpressure Protection
Adult
Child
Neonatal
Hardware and software double protections
<300 mmHg
<300 mmHg
<150 mmHg
Alarm indication
Sound and light alarming
Measurement Mode
Adult
Manual, Auto and STAT
Child
Manual, Auto and STAT
Neonatal
Manual, Auto
A.6.4 SpO
2
BLT SpO
2
SpO
2
Measurement range
0% to 100%
Measurement
technic
Digital SpO
2
technic
Resolution
1%
Accuracy
70% to 100%±2%
0%69%unspecified
Average time
Normal:8s, slow: 16sfast: 4s