Patient monitor User’s manual
17
Accuracy
20 bpm to 250 bpm: ±3 bpm
251 bpm to 300 bpm: unspecified
Resolution
1 bpm
Masimo SpO
2
SpO
2
Measurement range
0% to 100%
Resolution
1%
Accuracy
70% to 100% ±2% adult/pediatric, non-motion
conditions
70% to 100%±3%neonate, non-motion conditions
70% to 100%±3%motion conditions
0% to 69%unspecified
Average time
2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s
PR
Measurement range
25 bpm to 240 bpm
Accuracy
±3 bpmnon-motion conditions
±5 bpmmotion conditions
Resolution
1 bpm
SpCO
Measurement range
0% to 100%
Accuracy
0% to 40%±3%non-motion conditions
40%unspecified
SpMet
Measurement range
0% to 100%
Accuracy
0% to 15%±1%non-motion conditions
15%unspecified
PI
Measurement range
0.05% to 20%
SpHb
Measurement range
0 g/dl to 25 g/dl
Accuracy
8 g/dl to 17 g/dl±1g/dlnon-motion conditions
8 g/dl or 17 g/dlunspecified