Patient monitor User’s manual
24
Alarm range
SEV ISOENF
HALDES
0% to 30%, high/low limit can be adjusted continuously.
CO
2
0% to 10%, high/low limit can be adjusted continuously.
N
2
O
0% to 100%, high/low limit can be adjusted continuously.
Note1: The accuracy specification is valid for the operating temperature and humidity
conditions specified, except for interference specified in the table Interfering gas and
vapor effects below.
The accuracy specification for IRMA AX+ is not valid if more than two agents are present
in the gas mixture.
Interfering gas and vapor effects
If the concentration of the various gases is below the following level, there
will be negligible interference for measurement under the specification above.
IRMA CO
2
sensor:
Gas or vapor
Gas level
Effect (CO
2
)
Effect
(Agents)
Effect
(N
2
O)
N
2
O
60%
HAL
4%
ENF, ISO, SEV
5%
+8% of reading
1)
DES
15%
+12% of reading
1)
Xe (Xenon)
80%
-10% of reading
1)
He (Helium)
50%
-6% of reading
1)
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol)
0.3%
C
3
H
7
OH
(Isopropanol)
0.5%
CH
3
COCH
3
(Acetone)
1%
CH
4
(Methane)
3%
CO (Carbon
monoxide)
1%
NO (Nitrogen
monoxide)
0.02%
O
2
100%