EasyManua.ls Logo

Anyview A6 - Page 292

Anyview A6
305 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient monitor Users manual
53
Sensor Over Temp
Temperature of the sensor is over the
normal working temperature.
Low
Check CO
2
Sampling Line
Sampling tube is occluded or
damaged;
Sampling tube is kinked or pinched
Exhaust tube is blocked.
Zero Required
The sensor needs verify zero
Sensor no initialized
Sensor or module is not initialized
Prompt messages:
Messages
Cause
Level
Zero in Progress
Zeroing is in progress.
No alarm
Sensor Warm Up
Module is warming up.
AG
Physiological alarm:
Alarm messages
Cause
Level
EtAG labelhigh
Et AG label is above high alarm
limit
Low
User-Selectable
EtAG labellow
EtAG labelis below low alarm limit
FiAG labelhigh
Fi AG label is above high alarm
limit
Medium
User can select
Medium or
High
FiAG labellow
FiAG label is below low alarm
limit
Low
User-Selectable
EtN
2
O high
EtN
2
O is above high alarm limit
EtN
2
O low
EtN
2
O is below low alarm limit
FiN
2
O high
FiN
2
O is above high alarm limit
Medium
User-Selectable
FiN
2
O low
FiN
2
O is below low alarm limit
Low
User-Selectable
EtO
2
high
EtO
2
is above high alarm limit
Medium
User-Selectable
EtO
2
low
EtO
2
is below low alarm limit
FiO
2
high
FiO
2
is above high alarm limit
FiO
2
low
FiO
2
is below low alarm limit
Medium
User can select
Medium or
High
FiO
2
< 18%
FiO
2
measurement value is below
18%
High
Technical alarm:

Table of Contents

Related product manuals